TECHNICAL SPECIFICATIONS.
IMAGES | 2D | 3D |
Type | Complete examination of the 2 arches in a single scan for adults and children (reduced collimation). Studies of the maxillary region with maxillary sinuses. Studies localised to region of interest. |
Complete examination of the 2 arches in a single scan for adults and children (reduced collimation). Studies of the maxillary region with maxillary sinuses. Studies localised to region of interest. |
(Maximum) theoretical resolution on the patient plane |
2D: 5 – 6.9 lp/mm (pixel 100-73 µm) CEPH: 5.6 lp/mm (pixel 89 µm) |
≥ 6 LP/mm (Voxel 80 µm minimum section thickness) |
Equivalent radiograph size (cm) | PAN: 26.2 (length) x 14.4 (height) – CEPH: 29.2 (length) x 22 (height) |
|
Field of view (mm) | PAN: 210 (length) x 115 (height) CEPH: 258 (length) x 194 (height) PAN Child: 180 (length) x 100 (height) Dentition: 140 (length) x 100 (height) Bitewing: 167 (length) x 70 (height) |
DENT and SIN: 100 (diameter) x 100 (height) 100 (diameter) x 70 (height) 100 (diameter) x 60 (height) 80 (diameter) x 70 (height) 80 (diameter) x 60 (height)80 (diameter) x 100 (height) 60 (diameter) x 70 (height) 60 (diameter) x 60 (height) |
Maximum image data dimensions |
PAN: 7.5 MB (single image) CEPH: 14 MB |
720 MB |
Magnification | PAN: 1.2 – 1.3 CEPH: 1.13 |
1 to 1 (Isotropic voxel) |
Scan time | PAN: 13.7 s (ORTHO), 12.2 s (STD), 6.8 s (ECO)
CEPH: 9.9 s (REG); 3.7 s (ECO)
|
BEST QUALITY: 16.8 s (High Resolution) REGULAR QUALITY: 9.6 s (Standard)ECO QUALITY: 6.4 s (Low Dose) |
Estimate of typical effective dose (ICRP 103) |
PAN: 5 – 9 µSv | FOV: 10×10 | 35 µSv (Voxel 160 µm) – 121 µSv (Voxel 80 µm) FOV: 6×6 | 9 µSv (Voxel 160 µm) – 40µSv (Voxel 80 µm) |
Minimum image display times | RealTime | 15 s |
Advanced filters | ApT (Autoadaptive Picture Treatments) Self-adaptive PAN |
aMAR (Autoadaptive Metal Artifact Removal) |
*Optional vertical collimation on 2D PAN version (included in the version “CEPH Ready” and 3D) |
VERSION | 2D PAN o 2D PAN “CEPH READY” & 2D PAN-CEPH | 3D PAN “CEPH READY” & 3D PAN-CEPH |
Focal spot | 0.5 mm (IEC 60336) | 0.6 mm (IEC 60336) |
Tensione anodica | 60 kV – 85 kV continuous emission 60-70 kV continuous emission (PAN version only) |
2D mode: 60 kV – 85 kV continuous emission 3D mode: 90 kV (pulsed mode) |
Inherent filtration | > 2.5 mm Al eq. (at 85 kV) | 3D: 6 mm Al eq. (at 90 kV) – with automatic 3.5 mm filter |
Generator type | Constant potential (DC) |
Anode current | 4 mA – 15 mA |
Exposure control | 42 W (1:20 at 85 kV/10 mA) |
Maximum continuous anode input power |
42 W (1:20 at 85 kV/10 mA) |
Embedded X-ray shielding behind receptor |
IEC60601-1-3 compliant |
DETECTOR |
FUNCTION | PAN or PAN-CEPH | 3D and PAN | CEPH |
Type | CMOS (CsI) | Amorphous Silicon (CSI) | CMOS (CsI) |
Dynamic range | 14 bit (16,384 grey levels) | 16 bit (65,536 grey levels) | 14 bit (16,384 grey levels) |
ERGONOMICS
Examination selection | Procedure guided from virtual control panel on PC and/or iPad |
Patient positioning | Suggestion from virtual control panel – Servo-assisted alignment, 3 laser guides (Class 1 – IEC 60825-1) – 3D Scout View |
Patient positioning | Efficient 4 contact point 2D version – 5 contact point version, adjustable 3D/2D right/left |
Adjustments | 2-speed height adjustment drive Keypad on machine and/or iPad app Servo-assisted alignment: Keypad on machine or remote control (via Scout View) |
Other functions | Multilingual, parking position, remote control |
Notes | Easy access for patients in wheelchairs |
CONNECTIVITY
CONNECTIVITY | LAN / Ethernet |
Software | NewTom NNT (ISDP©10003:2018 compliant in accordance with EN ISO/IEC 17065:2012 certificate number 2019003109-1) and iPad App – NNT viewer (free), STL (RealGUIDE) |
Supported protocols | DICOM 3.0, TWAIN, VDDS, CLOUD shared (RealGUIDE) |
DICOM nodes | IHE compliant (Print; Storage Commitment; WorkList MPPS; Query Retrieve) |
iPad App | Virtual control panel for the device and for the NNT 2D viewer |
INSTALLATION
Minimum space requirement (L x D) | CEPH Ready version: 872 x 1030 mm CEPH version: 1785 x 1030 mm |
Package dimensions (L) x (D) x (H) in mm |
Box1: 930 x 690 x 960 (base machine) Box2: 1860 x 355 x 350 (floor-mounted) Box3: 575 x 1275 x 380 (teleradiographic arm) |
Weight | 2D version: 87 kg (192 lb) 3D/2D version: 99 kg (218 lb) CEPH option: 21 kg (46 lb) |
Accessories | Free standing base |
POWER SUPPLY
Voltage | Frequency | 115 – 240 V single-phase | 50 / 60 Hz |
115 – 240 V single-phase | 50 / 60 Hz | 20 A at 115 V; 12 A at 240 V |
20 A at 115 V; 12 A at 240 V | Maximum 0,5 A (240 V); 1 A (115 V) |
Notes | Automatic adaptation for voltage and frequency |
High quality images that meet a wide range of clinical diagnostic needs, all in one compact device.
GO 2D/3D/CEPH is a flexible platform that comes ready for the optional integration of the teleradiographic arm in a 2D or 3D configuration. Able to provide high resolution images, the platform prioritises patient health thanks to low exposure protocols and exclusive SafeBeamTM technology, which lets users adapt the dose to their actual diagnostic needs and the size of the scanned anatomical area. Excellent ergonomics and an adaptive alignment system ensure correct positioning of the patient and perfect focusing for clear, detailed images. A virtual control panel guides the operator through each stage of the examination. NNT is the technologically advanced software platform to manage, process, consult and share diagnostic images.