TECHNICAL SPECIFICATIONS.
2D images |
2D version | PAN | CEPH |
Main Examinations | • Panoramic Multilayer • Quadrants, • Bitewing • Maxillary Sinuses (AP and LL) • TMJ PA-LL |
Adds, with respect to the PAN version, teleradiography • Latero-Lateral • Antero-Posterior • Sub Mento Vertex • Carpus |
Child examination | Yes | Yes |
Maximum resolution | 6.3 – 7.5 lp/mm (Pixel 70-80 µm) | 5.6 lp/mm (Pixel 90 µm) |
Reduced fields of view (cm) | Length x Height • 22 x 12.9 (Child PAN); • 17 x 11 (Complete DENT)• 6 x 9 (BITEWING) |
Length x Height • 20 x 22 (Adult) • 30 x 18 (Child)• 20 x 18 (Child) |
3D eXtra Function | – | – |
Maximum image data size | 8 MB | 14 MB |
Magnification factor | PAN 1.25 (constant) | 1.13 |
ECO Scan scan time | Adult: 6 s | Child: 5.6 s | Low Adult: 4.5 s | Child: 3.3 s |
Standard scan time | Adult: 12.3 s | Child: 11.2 s | Complete Adult: 9 s |
Image display times | Real-Time | |
Advanced filters | ApT (Autoadaptive picture Treatments) | |
FULL-TOUCH 10” console on-board | Optional |
3D images |
3D version | PRIME | ADVANCED | PROFESSIONAL |
Main Examinations | Compared to the 2D version, it features 3D analysis of: • 2 dental arches in a single scan for adults and children with reduced collimation; • maxillary region with maxillary sinuses; • studies localised to DENTAL region of interest or to single TMJ. |
Compared to the PRIME version, it features 3D analysis of: • upper airways, either complete or partial, with variable collimation for frontal sinuses, nose and throat; • zygomatic implants; • one internal ear; • localised study of few teeth with maximum collimation or maximum effective resolution for endodonticexaminations or to evaluate micro- fractures. |
Compared to the ADVANCED version, it features 3D analysis of: • the whole Dental-Maxillofacial region; • both ears; • panoramic view with two temporomandibular joints; • cervical column. |
Child examination | Yes | Yes | Yes |
Resolution | Voxel 75 – 300 μm | Voxel 68 – 300 μm | Voxel 68 – 300 μm |
Maximum field of view (cm) | 10 (diameter); 8 (height) | 13 (diameter); 16 (height) | 16 (diameter); 18 (height) |
Available fields of view FOV Diameter x Height (cm) |
• 10 x 8; 10 x 6; 8 x 8; 8 x 6; 6 x 6 | • 13 x 16; 13 x 14; 13 x 10; 13 x 8; 10 x 10; • 10 x 8; 10 x 6; 8 x 8; 8 x 6; 6 x 6 |
• 16 x 18; 16 x 10; 15 x 6; • 13 x 16; 13 x 14; 13 x 10; 13 x 8; 10 x 10; • 10 x 8; 10 x 6; 8 x 8; 8 x 6; 6 x 6 |
3D eXtra Functions FOV Diameter x Height (cm) |
NO | 7 x 6; 4 x 4 | 9 x 16; 7 x 6; 4 x 4 |
Maximum image data size | < 495 MB | 215 MB – 820 MB | 360 MB – 820 MB |
ECO Scan scan time (exposure time) | 3.6 – 6.4 s (0.9 – 1.6 s) |
6.4 s – 26 s (1.6 s – 4.8 s) |
6.4 s – 26 s (1.6 s – 4.8 s) |
Regular Mode scan time (exposure time) |
14.4 s (3.6 s) |
14.4 s – 28.8 s (3.6 s – 7.2 s) |
14.4 s – 28.8 s (3.6 s – 7.2 s) |
Best Quality scan time (exposure time) |
26.4 s (8 s) |
16.8 s – 33.6 s (5.2 s – 10.4 s) |
16.8 s – 33.6 s (5.2 s – 10.4 s) |
Mean image viewing time | Minimum: 15 s | Minimum: 1 s | Minimum: 1 s |
Minimum: 1 s | aMAR (Auto-adaptive Metal Artifact Reduction) |
aMAR (Auto-adaptive Metal Artifact Reduction) | Supplied, except for the PRIME version (optional) |
X-ray generator
Generator type | Constant potential (DC) – High frequency 100-180 kHz | ||
Anode voltage | 2D: 60 kV – 85 kV – 3D: 90 kV (Pulsed mode) | ||
Anode current | 2 mA – 16 mA | ||
Focal spot | 0.5 mm (IEC 60336) – Fixed anode | ||
Exposure Control | Auto-Adaptative with intensity modulation during rotation –SafeBeamTM Technology | ||
Maximum continuous anode input power | 42 W (1:20 at 85 kV/10 mA) | ||
Inherent filtration | 2D: >2.5 mm Al eq. (at 85kV)-3D: 6.5 mm Al eq. (at 90 kV) |
Detector | |||
Detector type | 2D: CMOS – 3D: Amorphous Silicon (CsI) | ||
Dynamic range | 2D: 14 bit (16383 grey levels) – 3D: 16 bit (65536 grey levels) |
Ergonomics | |||
Patient alignment | 4 laser guides | ||
Patient positioning | 7 head contact points | ||
Adjustments | On-board keypad and/or virtual console for iPad (2-speed height drive) | ||
Examination selection | Virtual console on PC, Windows tablet and/or iPad and from Full-Touch 10” on-board console | ||
Notes | Easy access for patients in wheelchairs |
Connectivity
Connections | LAN / Ethernet | |
Software | NNT with Viewer software, free | |
Supported protocols | DICOM 3.0, TWAIN, VDDS | |
DICOM nodes | IHE compliant (Print; Storage Commitment; WorkList; MPPS; Query Retrieve) | |
App iPad | Virtual control panel for the device and for the NNT 2D viewer |
Installation
Minimum available work space requirement | 2D and 3D PAN: 1390 x 1140 mm – 2D and 3D CEPH: 1390 x 1800 mm |
Package dimensions (L) x (D) x (H) in mm | Machine Base: 1515 x 1750 x 670 mm – CEPH application: 1030 x 530 x 360 mm |
Weight | 2D PAN: 155 Kg – 342 lbs 2D CEPH: 175 Kg – 386 lbs 3D PAN: 155 Kg – 342 lbs 3D CEPH: 175 Kg – 386 lbs |
Accessories | Wall bracket even at 45° or floor support, free standing base available User-friendly for patients on wheelchair |
Power supply
Voltage | Frequency | 115 – 240 Vac, +/- 10% | 50/60 Hz +/- 2 Hz |
Maximum absorbed surge current | 20 A at 115 V; 12 A at 240 V |
Absorbed power in stand-by mode | 25 Watt |
Notes | Automatic adaptation for voltage and frequency |
SKU: NewTom GiANO HR
Description
GiANO HR is NewTom’s versatile and updatable device for all radiology-related needs. With a complete range of 2D and 3D examination options for dentistry, it offers specific multiple volumetric examinations for maxillofacial surgery, otorhinolaryngology and cervical spine examinations. Only top quality imaging with NewTom’s technology and experience.
Reviews (0)
Reviews
There are no reviews yet.